The device question

Is your software a medical device? Every other HDSF answer depends on this one. This guide helps you ask it well and record the answer. It does not answer it for you: the decision is the manufacturer's, under the MDR or IVDR, and it often needs a regulatory expert or a notified body.

The key idea: intended purpose

Qualification depends almost entirely on the intended purpose you claim, not on how the software works (MDCG 2019-11 rev. 1, June 2025). The same algorithm can be wellness in one app and a medical device in another, because of what each app says it does.

So write the intended purpose first (CORE-QUA-01), and make every text match it (CORE-QUA-03): store listings, website, onboarding, notifications, AI answers.

The questions, in order

Follow the decision trees in MDCG 2019-11 rev. 1 and record your answer to each step.

  1. Is it software? Not just data, a document or a database used as such.
  2. Does it drive or influence a hardware medical device? Then it follows that device.
  3. Does it do more than store, archive, communicate or simply search data? Software that only stores or sends data, without changing it for a medical result, is usually not a device.
  4. Is it for individual patients? Software for population statistics or general knowledge usually is not a device.
  5. Does it have a medical purpose in MDR Article 2(1)? Diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease; or of an injury or disability; and some other listed purposes.
  6. Does it analyse data from samples (in vitro)? Then the IVDR applies instead of the MDR.
  7. Is it modular? Some modules may be devices and others not. Mark the boundary clearly.
  8. Does it contain AI? It qualifies the same way; if it is a device, the AI Act adds requirements (MDCG 2025-6).

Signs you have a medical purpose

Signs you probably do not

These are signs, not answers. Disease prevention software can fall under Rule 11; read the MDCG examples closely.

If it is a device

14971), clinical evaluation (MDCG 2020-1) and post-market surveillance.

If it is not

(W-QUA-02).

(CORE-QUA-04).

Record it

Copy templates/qualification.md to docs/qualification.md, answer each step, sign and date it, and set in hdsf.yaml:

qualification:
  medical_device: none      # none | mdsw | ivd | annex_xvi
  class: null
  rationale: docs/qualification.md

"Undecided" is a valid state while you build, but not a launch state: the scanner blocks it.

Watch

The Commission proposed a targeted revision of the MDR and IVDR on 16 December 2025, which changes software classification rules (MDR-REV). HDSF will update this guide when the text is adopted.