The device question
Is your software a medical device? Every other HDSF answer depends on this one. This guide helps you ask it well and record the answer. It does not answer it for you: the decision is the manufacturer's, under the MDR or IVDR, and it often needs a regulatory expert or a notified body.
The key idea: intended purpose
Qualification depends almost entirely on the intended purpose you claim, not on how the software works (MDCG 2019-11 rev. 1, June 2025). The same algorithm can be wellness in one app and a medical device in another, because of what each app says it does.
So write the intended purpose first (CORE-QUA-01), and make every text match it (CORE-QUA-03): store listings, website, onboarding, notifications, AI answers.
The questions, in order
Follow the decision trees in MDCG 2019-11 rev. 1 and record your answer to each step.
- Is it software? Not just data, a document or a database used as such.
- Does it drive or influence a hardware medical device? Then it follows that device.
- Does it do more than store, archive, communicate or simply search data? Software that only stores or sends data, without changing it for a medical result, is usually not a device.
- Is it for individual patients? Software for population statistics or general knowledge usually is not a device.
- Does it have a medical purpose in MDR Article 2(1)? Diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease; or of an injury or disability; and some other listed purposes.
- Does it analyse data from samples (in vitro)? Then the IVDR applies instead of the MDR.
- Is it modular? Some modules may be devices and others not. Mark the boundary clearly.
- Does it contain AI? It qualifies the same way; if it is a device, the AI Act adds requirements (MDCG 2025-6).
Signs you have a medical purpose
- It suggests a diagnosis, or a likely condition, for one person.
- It calculates a dose, or tells someone to change a treatment.
- It monitors a disease and alerts when values cross clinical thresholds.
- It predicts a disease risk for one person and suggests what to do.
- It claims to detect, treat, cure or prevent a disease.
Signs you probably do not
- General fitness, sleep, diet or wellbeing tips, for people in general.
- Logging and showing data the person enters, without medical interpretation.
- Booking, payment, messaging or record keeping without changing the data for a medical result.
These are signs, not answers. Disease prevention software can fall under Rule 11; read the MDCG examples closely.
If it is a device
- Find the class (MDR Annex VIII, Rule 11 for most software; IVDR Annex VIII).
- Set up the quality management system (ISO 13485) and software life cycle (IEC 62304), risk management (ISO
14971), clinical evaluation (MDCG 2020-1) and post-market surveillance.
- Above class I (MDR) or class A (IVDR), you need a notified body.
- In app stores, show the device information before download, and list the app as a medical device (MDCG 2025-4).
- Profile M applies. Add
Mtoprofilesinhdsf.yaml.
If it is not
- Profile W usually applies. Keep the product inside its purpose (W-QUA-01) and its texts free of medical claims
(W-QUA-02).
- Check the qualification again at each release that adds a feature, a claim, an AI model or a user group
(CORE-QUA-04).
Record it
Copy templates/qualification.md to docs/qualification.md, answer each step, sign and date it, and set in hdsf.yaml:
qualification:
medical_device: none # none | mdsw | ivd | annex_xvi
class: null
rationale: docs/qualification.md
"Undecided" is a valid state while you build, but not a launch state: the scanner blocks it.
Watch
The Commission proposed a targeted revision of the MDR and IVDR on 16 December 2025, which changes software classification rules (MDR-REV). HDSF will update this guide when the text is adopted.